✉ info@eieinstruments.com
Home/Shop by Category/Civil Engineering Testing Instruments/Rock Testing Instruments
Hoek Cell For Rock Triaxial - Nx - 54.7 Mm

Hoek Cell For Rock Triaxial - Nx - 54.7 Mm

Hoek cells have been designed to be used for triaxial testing of rock specimens. Hoek cells comprise a steel body complete with two quick release self-sealing couplings, two steel end caps which are screwed to the cell body, 2 pieces of upper and 2 pieces of lower loading caps with spherical coupling and a rubber sealing sleeve to separate the specimen from the cell fluid.

Compliance:
  • ASTM D7012,IS 13047:1991,IS 9179,ASTM D5407,EN 1926 / EN 14580, ISRM
Applications:

Hoek Cells are designed for Testing Triaxial strengths of Rock or Concrete samples up to a Pressure of 40 MPa. These are designed to accept samples of NX size. Tests carried out on a series of samples under different confining pressures allow the user to determine:

  • Strength & Elastic properties
  • Shear strength at different confining pressures
  • Angle of Shearing Resistance & Cohesion
  • Modulus of Elasticity & Poisson’s Ratio
Core Specifications:
  1. Sample Diameter: 54.7 mm (standard NX core).
  2. Maximum Lateral Pressure: 150 kg/cm² (150 bar / approx. 15 MPa).
  3. Body Finish: Hard chrome plated for rust-proofing and corrosion resistance.
  4. Inlet/Outlet: Equipped with two quick-release self-sealing couplings for efficient fluid management.
Key Components:
  • Cell Body: Robust steel construction designed to withstand high lateral pressures without deformation.
  • End Caps: Two threaded steel caps that screw directly onto the cell body to secure internal components.
  • Loading Caps & Pistons:
    • Upper Assembly: Two pieces featuring a spherical seat for uniform axial load distribution.
    • Lower Assembly: Two pieces featuring a radial groove for improved stability or fluid drainage.
  • Sealing Sleeve: One-piece rubber sleeve that isolates the rock specimen from the hydraulic oil, allowing for rapid core insertion and removal without draining the fluid.
  • Hoek Cell to accept samples of EX, AX, AQ, BX, BQ, NQ, HQ, HQ3, PQ, PQ3 also available.

Coming Soon...

Coming Soon...

Coming Soon...